 Manual Lensmeter of internall reading
This device is suitable to test the diopters of spherical lenses of
ophthalmic lens or cylindrical lens, diopter, axis(axle) of astigmatism
of cylindrical lens, diopter of the prism and diopter of the cornea -
contact lens
Characteristics :
The ocular one is provided with a unit of approach which range of
approach is of + 5D to be adapted to the different visions of the
different operators.
The retcle is divided with the scale of measurement and the dividing
disc of the degree of the prism.
The entire degree of the prism is read directly of the scale of
measurement and of the decimal scale estimated according to the
degree on the scale of measurement.
When prism diopter is great than 5^ , use of prism compensator is
necessary. There are 2 scale line on the prism compensator: the
upper line denotes angle, with the range of 0º - 180º, minimum
scale value of 5º; the lower line is the prism diopter with a range
of 15^ - 0 - 15^
Specifications :
Ranges of measurement of diopters:
0 to +25D
Minimal value of the scale:
0,125D at 0 to + 5D
0,25D at + 5D to + 25D
Axis of the astigmatism of the cylindrical lens:
0° - 180°
Diopter of the prism:
0 - 5 ∆
Angle bases of the prism:
0° - 180°
Regulator of the ocular visibility:
+ 5D
Size of the lens:
- 16mm a - 80mm
Total dimensions of the device:
275mm x 130mm x 455mm
Weight:
5,6 Kg
Power:
220V - 110V / 15W
|